Assessment of biomechanical properties of specially-designed miniplate patterns in a mandibular subcondylar fracture model with finite element analysis and a servohydraulic testing unit

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

The aim of our study was to compare the biomechanical stability of three different systems used to fix mandibular condylar fractures on synthetic polyurethane mandibles using a servohydraulic testing unit and finite element analysis (FEA). We measured the resistance to displacement loads causing 1.75 and 3.5 mm displacements of the fragments, and the maximum resistance values before failure of the fixation system in 30 condylar necks of 15 polyurethane mandible models. The three subgroups investigated were arranged as a triangular plate (n = 10), a quadrangular plate (n = 10), and a straight miniplate (n = 10). The most successful fixation pattern was achieved with two straight four-hole miniplates. However, the triangular plate can easily be applied as a second option by an endoscopic approach. The rectangular plates showed no success in mandibular condylar fractures.

Original languageEnglish
Pages (from-to)848-853
Number of pages6
JournalBritish Journal of Oral and Maxillofacial Surgery
Volume58
Issue number7
DOIs
Publication statusPublished - Sept 2020

Keywords

  • biomechanics
  • finite element analysis
  • mandibular condyle
  • surgical fixation

Fingerprint

Dive into the research topics of 'Assessment of biomechanical properties of specially-designed miniplate patterns in a mandibular subcondylar fracture model with finite element analysis and a servohydraulic testing unit'. Together they form a unique fingerprint.

Cite this