Comparison and validation of finite element analysis with a servo-hydraulic testing unit for a biodegradable fixation system in a rabbit model

O. Atali, A. Varol, S. Basa, C. Ergun, S. Hartomacioǧlu

Research output: Contribution to journalArticlepeer-review

6 Citations (Scopus)

Abstract

The aim of this study was the biomechanical validation of three-dimensional finite element analysis (FEA) with a servo-hydraulic testing unit (STU) for a resorbable fixation system (RFS) in a rabbit model. Bilateral mandibular vertical body osteotomies (BMVBO) were performed in 15 female New Zealand rabbits. The animals were divided into three groups. The STU and FEA tests were done immediately after surgery in group 1 (1 day), at the first postoperative month in group 2, and at the third postoperative month in group 3. Both stress tests were carried out by applying vertical forces at the lower incisal edge, loading from 0 N force and increasing this until breakage occurred at the bone. The maximum forces that the hemimandibles could stand and the amount of deformation were recorded and analysed with the FEA and STU tests. We found the STU and FEA test results to be similar and that they could be used interchangeably for groups 1 and 3. However, the FEA results differed most from the real STU values in group 2 because of callus formation that had not ossified at the osteotomy line.

Original languageEnglish
Pages (from-to)32-39
Number of pages8
JournalInternational Journal of Oral and Maxillofacial Surgery
Volume43
Issue number1
DOIs
Publication statusPublished - Jan 2014
Externally publishedYes

Keywords

  • biodegradable implant
  • biomechanics
  • finite element analysis
  • mandible
  • osteotomy

Fingerprint

Dive into the research topics of 'Comparison and validation of finite element analysis with a servo-hydraulic testing unit for a biodegradable fixation system in a rabbit model'. Together they form a unique fingerprint.

Cite this