Mikro denetleyici tabanli biyoempedans analiz cihazi

Translated title of the contribution: Microcontroller based bioimpedance analyzer

Hakan Solmaz, Yekta Ülgen, Murat Tümer

Research output: Chapter in Book/Report/Conference proceedingConference contributionpeer-review

Abstract

The device presented in this study is a four-probe, multi-frequency, portable bioimpedance analyzer based on the magnitude-ratio and phase-difference detection principles. The system is built around a digital sine-wave generator, a voltage controlled current source, a phase-gain detector and a microcontroller unit. The Cole-Cole parameters, namely R0, R∞, fc and α are also calculated. The accuracy and precision of the device are checked against the HP 4284A LCR meter using different RC test loads.

Translated title of the contributionMicrocontroller based bioimpedance analyzer
Original languageTurkish
Title of host publicationProceedings of 2009 14th National Biomedical Engineering Meeting, BIYOMUT 2009
DOIs
Publication statusPublished - 2009
Externally publishedYes
Event2009 14th National Biomedical Engineering Meeting, BIYOMUT 2009 - Balcova, Izmir, Turkey
Duration: 20 May 200922 May 2009

Publication series

NameProceedings of 2009 14th National Biomedical Engineering Meeting, BIYOMUT 2009

Conference

Conference2009 14th National Biomedical Engineering Meeting, BIYOMUT 2009
Country/TerritoryTurkey
CityBalcova, Izmir
Period20/05/0922/05/09

Fingerprint

Dive into the research topics of 'Microcontroller based bioimpedance analyzer'. Together they form a unique fingerprint.

Cite this